Detailed troubleshooting tips to help you resolve flow cytometry issues for cell cycle analysis purpose. Though the troubleshooting tips provided here cover many different problems you may encounter, we do not expect it to be the exclusive solution to any problems. We hope that you will find the information beneficial to you and useful as a reference guide in troubleshooting cell cycle staining flow cytometry problems. If you need more assistance with your experiments, please contact us for additional help.
Click for the Protocol of Cell Cycle Staining Flow Cytometry.
Contents
No Signal / Weak Fluorescence Intensity
Cause | Solution |
Insufficient detection antibody | Increase the concentration of detection antibody. |
Sub-optimal antigen-antibody binding |
|
The surface antigen is getting internalized | Perform all steps at 4°C and use ice cold reagents. |
Target protein not present or expressed at low level | Ensure tissue or cell type expresses target protein and that it is present high enough to be detected. |
Antibody storage |
|
Higher cell concentration | Adjust cell population to recommended density. |
Inadequate fixation and/or permeabilization of sample. | 70% ethanol (ice-cold) can sometimes be used as an alternative method of fixation, particularly when performing cell cycle analysis. |
Others |
|
High Fluorescence Intensity
Cause | Solution |
High antibody concentration |
|
Inadequate blocking | Dilute antibodies in blocking solution. |
A high expressing antigen is paired with bright fluorochrome | Always pair strong antigens with dimmer fluorochromes such as FITC or Pacific Blue. |
The fluorescent signal is under-compensated | Use MFI alignment instead of visual comparison to compensate. |
High Background
Cause | Solution |
The gain set is too high, the offset is too low |
|
Excess antibody |
|
Non-Specific Staining
Cause | Solution |
High autofluorescence |
|
Non-specific cells are targeted |
|
Excess, unbound antibodies are present in the sample | Wash cells adequately after every antibody incubation step. |
Presence of dead cells |
|
Low Event Rate
Cause | Solution |
Low number of cells | Optimal cell concentration for each sample should be around 1x106 cells/mL. |
Cells clumped/blocking tubing | Pipet the samples gently before staining and again before running the cytometer. |
High Event Rate
Cause | Solution |
High number of cells | Optimal cell concentration for each sample should be around 1x106 cells/mL. |
Contamination | Repeat the staining procedures. |
Unusual Scatter Profiles
Cause | Solution |
The cells are lysed or damaged |
|
Bacterial contamination |
|
Presence of dead cells |
|
Incorrect instrument settings for scatter |
|
Loss of Epitope
Cause | Solution |
Sample was not kept on ice | Keep antibodies at 4°C to prevent loss of activity. This also prevents active phosphatases and proteases from altering the epitope of interest. |
Two or More Cell Populations Are observed When There Should Be Only One Population
Cause | Solution |
Cell doublets are present in the sample |
|
There is more than one cell population present which is expressing the target protein | Check the expected expression levels from the cell types that are present in the sample and ensure adequate cell separation if necessary. |
When running samples to examine cell cycle distribution the histogram for DNA content does not resolve the distinct phases of the cell cycle, i.e., G0/G1, S and G2/M phases.
Cause | Solution |
Incorrect flow rate | Ensure that your samples are being run at the lowest flow rate setting on your cytometer. High flow rates will give rise to high coefficients of variation (CVs), leading to a loss of resolution of the different phases of the cell cycle. |
Insufficient staining with Propidium Iodide/RNase (PI) solution. | Resuspend cell pellet directly in PI/RNase solution and incubate for at least 10 min. If PI cannot be used due to fluorescent channel configurations, change an alternate DNA dye. |
Cells are not proliferating. | Cells should be harvested during asynchronous and exponential growth to ensure that all phases of the cell cycle are accurately represented. |
Other Possible Problems
Cause | Solution |
Population shifts during measurement | Too much PI in your stained sample. Decrease the amount of PI. |
Population shifts between samples | Cell numbers are different. Optimal cell concentration for each sample should be around 1x106 cells/mL. |
Large CVs and weak resolution of histograms | Too much free DNA/ RNA in the cell suspension. |
Large CVs | The flow rate is to high. Optimize the flow rate. |
Tips