Immunofluorescence (IF) staining is a widely used technique in biological research and clinical diagnostics. Here Creative Biolabs lists the common IF troubleshooting tips. Though the tips provided here cover many different problems you may encounter in IF, we hope that you will find the information beneficial to you and useful as a reference guide. If you ever need more assistance with your experiments, please contact us for additional help.
Protocol of Immunofluorescence is available for you.
Contents
Weak or No Staining
Possible Cause | Recommended Solution |
Incorrect light source/filter set | Make sure your microscope is equipped with the correct light source/ filter set for the fluorophore you have chosen. |
Gain/exposure is too low | Turn up the gain and/or increase the exposure time. |
Fluorescent tag bleached | Avoid over exposure of the slide to light sources for extended periods. Always store slides in the dark. |
Samples are over fixed |
|
Cells are not permeabilized |
|
Tissue/cells dried out | Samples must be kept covered in liquid throughout the staining process. |
Not enough primary antibody | Use a higher concentration of antibody. |
The primary and secondary antibodies are incompatible |
|
Slide storage | Samples should be imaged shortly after processing as the signal decreases over time. Store slides at 4⁰C in the dark if needed. |
Antibody storage | Antibody was not stored as recommended. A new vial will need to be used if necessary. |
Low expression of protein of interest |
|
Incubation time is too short | Increase the incubation time. |
Inappropriate testing model | Protein expression should be confirmed by WB or other means. |
Target protein not induced properly | Optimal treatment conditions and controls should be determined for each target. |
Non-Specific Staining
Possible Cause | Recommended Solution |
Spectral overlap | Adjust your light sources/filters to pick up only one signal at a time. If this is not possible, choose new fluorophores that do not have spectral overlap. |
Antibody concentration is too high | Reduce the concentration, and the incubation period. |
Aggregates | Spin down secondary antibodies in a microcentrifuge to move aggregates to the bottom of the tube. |
The primary antibody is raised again the same species as the tissues stained |
|
High Background
Possible Cause | Recommended Solution |
Sample autofluorescence |
|
Insufficient blocking | Use normal serum from the same species as the secondary antibody used. |
Incorrect antibody dilution | Dilute according to recommended antibody dilution. |
Samples dried out | Samples must be kept covered in liquid throughout the staining process. |
Insufficient washing | Increase washing times. |
Antibody cross-reactivity | Use isotype control antibodies to determine whether your secondary antibody is cross-reacting. |
Non-specific antibody binding | Compare to knockdown or knockout cells or compare to cells known to express higher or lower levels of the target antigen. |
Negative Control Stains Positive and/or Positive Control Stains Negative
Possible Cause | Recommended Solution |
Use of wrong control or reagents | Check reagent labels and repeat test. |
Cross-contamination of controls and/or specimens |
|
Running of controls and/or specimens between wells | Use smaller drops to stay within well perimeters. |
Splashing of excess control and/or specimens from well to well due to over vigorous rinsing with wash bottle |
|
Cells Visible but Negative Staining with All Tests Including Positive Control
Possible Cause | Recommended Solution |
Reagents used in wrong sequence or one or more reagents not added | Review procedure and repeat test following staining steps precisely. |
More Positive Results than Expected (False Positives) with IgM Testing
Possible Cause | Recommended Solution |
Presence of rheumatoid factor | Pretreatment of serum, using ion exchange chromatography or IgG immunoprecipitation, to remove IgG interference. |
More Negative Results than Expected (False Negatives) with IgM Testing
Possible Cause | Recommended Solution |
Competition from IgG specific antibodies. | Pretreatment of serum, using ion exchange chromatography or IgG immunoprecipitation, to remove IgG interference. |
Weaker Results than Expected with IgM Testing
Possible Cause | Recommended Solution |
Incubation time too short | Increase the incubation time. |
Inadequate incubation temperature | Incubate at 37°C. |
Results Differ from Those Reported by Another Laboratory
Possible Cause | Recommended Solution |
Microscope optics, filters, light sources, reagents and personnel may vary enough to result in differences of twofold or more. | Each laboratory must establish their own normal ranges. Results between laboratories should not be compared. |
Notes for Fluorescent Double Staining
For Direct Method:
For Indirect Method:
The other procedures are the same as conventional IF experiments.